Pensky-Martens Flash Point tester (D93)
An exceptional time and cost saving instrument, the Petrotest® PMA 5 is the first and only 12-position Pensky-Martens flashpoint tester (single position shown). Featuring ASTM D93 (A&B&C), Search, Rapid, and user-customized test programs, the compact design enables the complete PMA 5 to be set up on the benchtop or in most lab fume hoods.
• Easy to operate
• Versatile
• Compact
• Precise
• Reliable and economical
Applications
Bitumen, Diesel Oil, Fuel, Lube Oil, Power Plant, Jet Fuel, Turbine Fuel, Engine Oil, Turbine Oil and Solvents
Method A:
for distillate fuels (diesel, kerosene, heating oil, turbine fuels), new lubricating oils, paints, vanishes and other homogeneous liquids.
Method B:
for residual fuel oils, cutback residual, used lubricating oils, non-homogeneous materials like mixtures of petroleum liquids & solids, surface-film building petroleum liquids or liquids with a viscosity above 5.5 cSt at +40°C. Used in shipping & safety regulations to define flammable and combustible materials.
Method C:
for biodiesel (FAME, B100), electronic flash point detection required. Used in shipping & safety regulations to define flammable and combustible materials.
• ASTM D93 A+B+C, Search, Rapid and User-Customized Test Programs
• Time and Cost Saving 12-Position Test Cup Autosampler Available
• 7 mL Test Cups Available Providing Equivalent Results to ASTM D93
• Temperature Range from 40 to 360°C
• Automatically Lights Test Flame, Relights if Necessary During Test, Suppresses Gas Source at end of Test
• Built-in Sensor for Automatic Barometric Pressure Correction
• Integrated Fire-Extinguishing System
Programs
4x ASTM Standards
4x ISO Standards
2x Rapid Heating
2x Search Run
2x User Defined
Ignition Type
gas and electric (included)
Stirring Speed
60 to 250 rpm (adjustable)
Supplied With
PMA 5 model, 1 cup, 1 cover, multi-detector, gas ignitor, electric ignitor, tongs for cup, try for cup and multi-detector, PC Software and power supply.
Technical Data
Temperature Range: +40 to +360 °C (+104 to +680 °F)
Ignition Type: gas and electric (included)
Stirring Speeds: 60 to 250 rpm (adjustable)
Sensing System: Differential-Thermocouple
Barometric Pressure Sensor: automatic correction of the measured values
Safety: overheat protection, automatic shut-off
Interface: RS-232 for printer, RS-232 for computer download
(USB-converter available)
Gas Connection: for propane/butane or natural gas (max. 0.05 bar)
Display: °C or °F (selectable)
Dimensions: Width 230 mm
Depth 470 mm, plus wall clearance
Height 460 mm
Weight 12 kg
Power Consumption: 900 Watts
Power Supply: 230/115 V, 50/60 Hz (selectable)
ASTM D93
Standard Test Method for Flashpoint by Pensky-Martens Closed Cup Tester
Standard(s):
Pensky-Martens Method: ASTM D93 A+B+C, JIS K2265, AASHTO T73, AASHTO T172, (EN 22 719), FTM 791-1102, FTM 141-4293, IP 34 A, B (BS 2839), ISO 2719 A, B (DIN 51758, NF M07-019)
Biodiesel Method: ASTM D6751
|
|
 |
 |
|
 |